Cytoflex4L
CytoFLEX S Blue-Red-Violet-Yellow (B-R-V-Y) has the capacity for 16 parameters, including 13 for fluorescence detection. The fully activated instrument includes two channels from the 488 nm (Blue) laser, three from the 638 nm (Red) laser, four from the 405 nm (Violet) laser, four from the 561 nm (Yellow Green). Expanded laser color palette provides flexibility to optimize panel design for efficient marker detection
Fixed integrated optics and quartz flow cell design enables portability
Avalanche Photodiode (APD) detectors provide unmatched red sensitivity and unique compensation workflow
Violet Side scatter (VSSC) optional configuration for nanoparticle resolution
Novel Compensation Library: store fluorescent spillover values of dyes to easily determine the correct compensation matrix with new gain settings
Absolute linear gain amplification enables the use of compensation settings between experiments and sample types
Equipped with CytExpert software for analysis